Fujifilm Irvine Scientific, Inc. recalls Sperm separation media are used to separate motile sperm from other constituents of semen.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The lower layer component of a sperm separation Kit does not match the certificate of analysis. The kit should contain lower layer (Lot: 0000020686, Shelf Life: 31 December 2024) but incorrectly may contain lower layer (Lot number 9925811106, Shelf Life: 30 November 2023).
- Products
-
- Sperm separation media are used to separate motile sperm from other constituents of semen.
- Company
- Fujifilm Irvine Scientific, Inc.
- Where it was sold
-
US: OK, CA, DC, NC, NV, TX, MI, IN, NY, UT, WI, MN, PA, WV, NJ, IL, LA, WA, OR, FL, IA. OUS: GB, NL, AR, MY
States: California, District of Columbia, Florida, Iowa, Illinois, Indiana, Louisiana, Michigan, Minnesota, North Carolina, New Jersey, Nevada, New York, Oklahoma, Oregon, Pennsylvania, Texas, Utah, Washington, Wisconsin, West Virginia
- Barcodes
00857515006009- Units affected
- 246
- About this recall
- Sperm separation media are used to separate motile sperm from other constituents of semen.