Brasseler USA I Lp recalls Brasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The device has a grip detail (right angle latch) out of specification causing the bur to not adequately latch in the handpiece.
- Products
-
- Brasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023
- Company
- Brasseler USA I Lp
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, IL, NY, OH, OR, PA, TN, TX, and WA. The country of Canada.
States: Alabama, California, Colorado, District of Columbia, Illinois, New York, Ohio, Oregon, Pennsylvania, Tennessee, Texas, Washington
- Barcodes
00887919188483- Units affected
- 440 burs
- About this recall
- Brasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023, REF 001156T0), carbide dental bur, 10 burs/package, Rx.