Beaver-Visitec International, Inc. recalls BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
- Products
-
- BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
- Where it was sold
-
US Nationwide distribution in the states of FL, MI, OK, TX, WI.
States: Florida, Michigan, Oklahoma, Texas, Wisconsin
- Barcodes
00886158037576- Units affected
- 5 units
- About this recall
- BVI Leos Laser and Endoscopy System. Model Number: OME6000U.