Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Boston Scientific Corporation recalls AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553650

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
Products
  • AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553650
Where it was sold

Nationwide including Puerto Rico Foreign: To be provided

Nationwide

Barcodes
08714729904595
Units affected
1144 units US; 60 units (OUS)
About this recall
AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553650