Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553520
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- Products
-
- HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553520
- Where it was sold
-
Nationwide including Puerto Rico Foreign: To be provided
Nationwide
- Barcodes
08714729904540- Units affected
- 263 units (OUS)
- About this recall
- HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553520