Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

OriGen Biomedical, Inc. recalls Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
Products
  • Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing
Company
OriGen Biomedical, Inc.
Where it was sold

US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE

States: California, Florida, Massachusetts, Maryland, North Dakota, Nebraska, New Jersey, Ohio, Texas, Washington

Units affected
3613 Cases
About this recall
Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T