Ventec Life Systems, Inc. recalls VOCSN Patient Breathing Package (Pediatric
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
- Products
-
- VOCSN Patient Breathing Package (Pediatric
- Company
- Ventec Life Systems, Inc.
- Where it was sold
-
U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan
States: California, Colorado, Florida, Hawaii, Indiana, Missouri, North Carolina, Washington
- Barcodes
00855573007594- Units affected
- 250 breathing circuits
- About this recall
- VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.