Maquet Medical Systems USA recalls Model No. BEQ-HLS 5050 USA
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.
- Products
-
- Model No. BEQ-HLS 5050 USA
- Company
- Maquet Medical Systems USA
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
04058863076355- Units affected
- 16520 worldwide; 6657 US
- About this recall
- Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with the CARDIOHELP heart-lung machine.