Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Maquet Medical Systems USA recalls CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.
Products
  • CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
Where it was sold

US Nationwide distribution in the states of NJ & TX.

States: New Jersey, Texas

Barcodes
04037691658384
Units affected
148 units (US: 5; OUS: 143)
About this recall
CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.