Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ethicon, LLC recalls ETHIBOND EXCEL Polyester Suture

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
Products
  • ETHIBOND EXCEL Polyester Suture
Where it was sold

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and the countries of AT, AU, BE, CA, CH, CN, CZ, DE, DK, FI, FR, GB, HU, IE, JP, LU, NL, NO, PL, SE, SG, ZA.

Nationwide

Units affected
48
About this recall
ETHIBOND EXCEL Polyester Suture, REF: 6664H, X425H, X872H, EH7491H. Not sold in US