Ethicon, Inc. recalls
Ethicon, Inc. is named on 9 recalls in Recall Relay's records since November 2023, from FDA (16 recalled products: FDA lists each product as its own recall); the most recent was September 23, 2026.
All 16 recalled products, newest first.
-
Class IIofficial
Ethicon, LLC recalls ETHILON Nylon SutureNylon suture indicated for use in general soft tissue approximation and/or ligation may have needles with distorted curvature, that may break during use, or could cause additional soft tissue or vessel injury, including tearing or bleeding intraoperatively.
-
Class IIofficial
Ethicon, LLC recalls PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical SutureSilk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
-
Class IIofficial
Ethicon, LLC recalls STRATAFIX Spiral PDS Plus Bidirectional Knotless Tissue Control DevicePotential for barb non-engagement.
-
Class IIofficial
Ethicon, LLC recalls PERMA-HAND Silk SutureSutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
-
Class IIofficial
Ethicon, LLC recalls ETHIBOND EXCEL Polyester SutureSutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
-
Class IIofficial
Ethicon, LLC recalls PROLENE Polypropylene SutureSutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
-
Class IIofficial
Ethicon, Inc. recalls MONOCRYL" (poliglecaprone 25) Suture REF Y936H MONOCRYL" Sutures are indicated for use in general soft tissueDue a component mix-up in manufacturing, the affected products may contain incorrect needle type/size and suture length.
-
Class IIofficial
Ethicon, Inc. recalls Coated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CTComplaints were received regarding needle pull-off and suture degradation. Analysis on returned product from this lot confirmed that some sutures were visually degraded in the package.
-
Class IIofficial
Ethicon, Inc. recalls MONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximationIssue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
-
Class IIofficial
Ethicon, Inc. recalls VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation anIssue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
-
Class IIofficial
Ethicon, Inc. recalls PDS II (polydioxanone) Suture - Intended for use in general soft tissue approximationIssue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
-
Class IIofficial
Ethicon, Inc. recalls VICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligationIssue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
-
Class IIofficial
Ethicon, Inc. recalls MONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation whIssue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
-
Class IIofficial
Ethicon, Inc. recalls PDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximationIssue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
-
Class IIofficial
Ethicon, Inc. recalls SURGIFOAM" Absorbable Gelatin Powder KitSome applicator tips from these lots had open seals.
-
Class IIofficial
Ethicon, Inc. recalls STRATAFIX Spiral MONOCRYL Plus Bidirectional Knotless Tissue Control Device: STRATAFIX SPIRAL MONOCRYL PLUS BIDegraded and unusable upon removal from the foil pouch due to exposed environmental conditions during transit and storage, compromising sterility
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.