Philips Medical Systems Nederland B.v. recalls Philips Azurion Interventional Fluoroscopic X-ray Systems
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
- Products
-
- Philips Azurion Interventional Fluoroscopic X-ray Systems
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
0088483805929000884838059207008848380852510088483809924100884838059191- Units affected
- 1,993 US; 2,662 ROW
- About this recall
- Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected PCs: FlexViewing PC Azurion Model Name - Model Number: Azurion 7 B12 722067; Azurion 7 B20 722068; Azurion 7 M12 722078 722223; Azurion 7 M20 722079;