Canon Medical System, USA, INC. recalls The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an inter
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.
- Products
-
- The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an inter
- Where it was sold
-
US: CA, LA
States: California, Louisiana
- Barcodes
04987670100192- Units affected
- 1
- About this recall
- The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.