Canon Medical System, USA, INC. recalls Alphenix INFX-8000H
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
- Products
-
- Alphenix INFX-8000H
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of Dominican Republic.
Nationwide
- Units affected
- 3 units
- About this recall
- Alphenix INFX-8000H, interventional fluoroscopic x-ray system