Ptw-freiburg recalls Software VERIQA (S070031)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.
- Products
-
- Software VERIQA (S070031)
- Company
- Ptw-freiburg
- Where it was sold
-
GA
States: Georgia
- Units affected
- 1 US; 65 worldwide
- About this recall
- Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.