Ptw-freiburg recalls Software VERIQA (S070031)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.
Products
  • Software VERIQA (S070031)
Where it was sold

GA

States: Georgia

Units affected
1 US; 65 worldwide
About this recall
Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.