Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ventec Life Systems, Inc. recalls VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
Products
  • VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro
Where it was sold

US distribution in CA, NY, and MO.

States: California, Missouri, New York

Units affected
30
About this recall
VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002