Tornier, Inc recalls TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
- Products
-
- TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE
- Company
- Tornier, Inc
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Units affected
- 427
- About this recall
- TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318