Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Tornier, Inc recalls Stryker Tornier Perform Humeral System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
Products
  • Stryker Tornier Perform Humeral System
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
00846832084494
Units affected
38 units
About this recall
Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty