Tornier, Inc recalls Stryker Tornier Perform Humeral System
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
- Products
-
- Stryker Tornier Perform Humeral System
- Company
- Tornier, Inc
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
00846832084494- Units affected
- 38 units
- About this recall
- Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty