Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Healthcare (China) Co., Ltd. recalls SIGNA Creator

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Products
  • SIGNA Creator
Where it was sold

Domestic US distribution nationwide. International distribution worldwide.

Nationwide

Barcodes
001952785772210019527857723800840682113762008406821137860084068211611400840682118576008406821189340084068211898900840682121729008406821468140019527812460900840682104401008406821185830084068211860600840682118958008406821189960084068213823900840682145770
Units affected
2538
About this recall
SIGNA Creator, SIGNA Explorer, NMRI systems