GE Healthcare (China) Co., Ltd. recalls SIGNA MR355

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Products
  • SIGNA MR355
Where it was sold

Domestic US distribution nationwide. International distribution worldwide.

Nationwide

Barcodes
008406821444070019527836125700840682144445
Units affected
241
About this recall
SIGNA MR355, SIGNA MR360, NMRI system