Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Focalyx Technologies, LLC. recalls Focalyx Fusion

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.
Products
  • Focalyx Fusion
Company
Focalyx Technologies, LLC.
Where it was sold

Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO.

States: Florida, Massachusetts, New York

Barcodes
00860003552001
Units affected
18
About this recall
Focalyx Fusion