Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Human Med Ag recalls human med Biofill Infiltration Cannula

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.
Products
  • human med Biofill Infiltration Cannula
Company
Human Med Ag
Where it was sold

US Nationwide distribution in the state of FL.

States: Florida

Barcodes
04260170881014
Units affected
775 devices
About this recall
human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.