Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Telcare, LLC recalls Philips Connected Blood Glucose Meter

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Telcare, LLC has received customer complaints related to battery expansion/swelling for its Philips Connected Blood Glucose Meter (i.e., BGM 4 brand name) Monitoring System.. The issue was identified by post market surveillance data (customer complaints) - events of battery expansion/swelling and associated symptoms, e.g., leading to the device's removable back cover (battery port) to separate. If there is a loss of primary function of the device due to the described device defect, an immediate health consequence that may result is prolonged hypoglycemia or hyperglycemia due to delay in treatment, as the blood glucose measurement is not available to guide treatment. In a worst-case situation, there is potential for battery swelling to result in explosion of the battery and/ or device. Exposure to explosion, and subsequently projectiles from the explosion, may result in several immediate health consequences, including concussion, headache, dizziness, eye irritation, blurred vision, eye injury, and vision impairment.
Products
  • Philips Connected Blood Glucose Meter
Company
Telcare, LLC
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
00859519002407
Units affected
34,764 units
About this recall
Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009;