Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

TAS Medical Inc recalls TISSUE APPROXIMATION SYSTEM (TAS)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Products
  • TISSUE APPROXIMATION SYSTEM (TAS)
Company
TAS Medical Inc
Where it was sold

US Nationwide distribution in the states of LA, OK, TN.

States: Louisiana, Oklahoma, Tennessee

Units affected
114
About this recall
TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP