TAS Medical Inc recalls TISSUE APPROXIMATION SYSTEM (TAS)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
- Products
-
- TISSUE APPROXIMATION SYSTEM (TAS)
- Company
- TAS Medical Inc
- Where it was sold
-
US Nationwide distribution in the states of LA, OK, TN.
States: Louisiana, Oklahoma, Tennessee
- Units affected
- 114
- About this recall
- TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP