FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
- Products
-
- FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23
- Where it was sold
-
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
States: Alaska, California, Connecticut, Florida, Illinois, New Hampshire, Ohio, Wisconsin
- Barcodes
045402170522260454021705743604540217057450- Units affected
- 16 units
- About this recall
- FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)