FUJIFILM Healthcare Americas Corporation recalls
FUJIFILM Healthcare Americas Corporation is named on 10 recalls in Recall Relay's records since November 2023, from FDA (16 recalled products: FDA lists each product as its own recall); the most recent was July 22, 2026.
All 16 recalled products, newest first.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls FUJIFILM Synapse PACS software v7.4.310; radiological image processing systemSoftware issue may cause the ruler measurement to be inaccurate.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls FDR Visionary SuiteIt was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls ASPRIRE Cristalle Mammography System (cleared under K212873) installed with the ASPIRE Cristalle Digital BreasDevices had an unapproved slabbing software function enabled for use.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000The incorrect computed patient age is showing in VX for patients less than 3 months old.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skullBolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS - Version 7.2.200Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS - Version 7.3.000Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS - Version 7.1.000USMeasurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS - Version 7.2.000Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS - Version 7.1.000Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS - Version 7.2.100Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical imaThe LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls Synapse Cardiology PACS V7.3.0The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
-
Class IIofficial
FUJIFILM Healthcare Americas Corporation recalls The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-IWhen the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.