Boston Scientific Corporation recalls Boston Scientific POLARSHEATH Steerable Sheath 12F
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Boston Scientific has identified a tooling error in manufacturing which may have caused delamination of the inner lumen of the sheath shaft in a subset of POLARSHEATH devices. This could result in the embolization of a fragment during use (such as flushing of the sheath or introduction of the dilator or ablation catheter).
- Products
-
- Boston Scientific POLARSHEATH Steerable Sheath 12F
- Where it was sold
-
Austria, Belarus, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Netherlands, Poland, Portugal, Spain, and Japan
- Barcodes
08714729992684- Units affected
- 461 devices
- About this recall
- Boston Scientific POLARSHEATH Steerable Sheath 12F, REF M004CRBS3050