Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Boston Scientific Corporation recalls Boston Scientific POLARSHEATH Steerable Sheath 12F

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Boston Scientific has identified a tooling error in manufacturing which may have caused delamination of the inner lumen of the sheath shaft in a subset of POLARSHEATH devices. This could result in the embolization of a fragment during use (such as flushing of the sheath or introduction of the dilator or ablation catheter).
Products
  • Boston Scientific POLARSHEATH Steerable Sheath 12F
Where it was sold

Austria, Belarus, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Netherlands, Poland, Portugal, Spain, and Japan

Barcodes
08714729992684
Units affected
461 devices
About this recall
Boston Scientific POLARSHEATH Steerable Sheath 12F, REF M004CRBS3050