Fujirebio Diagnostics, Inc. recalls Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
- Products
-
- Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable
- Company
- Fujirebio Diagnostics, Inc.
- Where it was sold
-
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
States: Arizona, California, Florida, Indiana, Kansas, Kentucky, Massachusetts, Maryland, Minnesota, New Jersey, Pennsylvania, Texas, Washington
- Barcodes
05414470815593- Units affected
- 179 units
- About this recall
- Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse ¿-Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing ¿-amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse ¿-Amyloid Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio