Preat Corp recalls Preat Nobel Active/Conical-compatible RP Titanium Blank
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.
- Products
-
- Preat Nobel Active/Conical-compatible RP Titanium Blank
- Company
- Preat Corp
- Where it was sold
-
U.S. Nationwide distribution in the states of AZ, OH and PA.
States: Arizona, Ohio, Pennsylvania
- Barcodes
00842092165867- Units affected
- 561 units
- About this recall
- Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967