Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Preat Corp recalls Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant.
Products
  • Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be
Company
Preat Corp
Where it was sold

U.S. Nationwide distribution in the states of AZ, CA, GA, MO, NJ, NY, OH, PA, RI, TN,TX, and WA.

States: Arizona, California, Georgia, Missouri, New Jersey, New York, Ohio, Pennsylvania, Rhode Island, Tennessee, Texas, Washington

Barcodes
00842092166093
Units affected
145 abutments
About this recall
Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.