Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Zimmer, Inc. recalls Z1 Femoral Hip System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is a potential design issue with the Z1 Offset Stem Inserter leading to fracturing of the distal tip.
Products
  • Z1 Femoral Hip System
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
00887868577888
Units affected
35 units
About this recall
Z1 Femoral Hip System, Product Number 611777612