Zimmer, Inc. recalls
Zimmer, Inc. is named on 14 recalls in Recall Relay's records since November 2023, from FDA (61 recalled products: FDA lists each product as its own recall); the most recent was May 20, 2026.
All 61 recalled products, newest first.
-
Class IIofficial
Zimmer, Inc. recalls Off-Axis Comprehensive Shoulder SystemFive complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
-
Class IIofficial
Zimmer, Inc. recalls Off-Axis Comprehensive Shoulder System SmallFive complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
-
Class IIofficial
Zimmer, Inc. recalls Off-Axis Alliance GlenoidFive complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
-
Class IIofficial
Zimmer, Inc. recalls Off-Axis Alliance GlenoidFive complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
-
Class IIofficial
Zimmer, Inc. recalls Off-Axis Comprehensive Shoulder SystemFive complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Persona Revision Product Name: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Model/CaFirm received a complaint with no clinical impact in which a posterior augment was found within the distal augment package.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 0Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/CatalAffected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
-
Class IIofficial
Zimmer, Inc. recalls Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked PolyethyleneZimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
-
Class IIofficial
Zimmer, Inc. recalls Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked PolyethyleneZimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
-
Class IIofficial
Zimmer, Inc. recalls Zimmer Biomet Ceramic HeadThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
-
Class IIofficial
Zimmer, Inc. recalls Zimmer Biomet Ceramic HeadThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
-
Class IIofficial
Zimmer, Inc. recalls Zimmer Biomet Ceramic HeadThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
-
Class IIofficial
Zimmer, Inc. recalls Zimmer Biomet Ceramic HeadThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
-
Class IIofficial
Zimmer, Inc. recalls Zimmer Biomet Ceramic HeadThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
-
Class IIofficial
Zimmer, Inc. recalls NexGen LPS FlexThe "Use with plate 7, 8, 9, 10" compatibility statement on top of the label does not match with the compatibility statements "USE WITH PLATE 5, 6" in the translations section of the same label.
-
Class IIofficial
Zimmer, Inc. recalls Z1 Femoral Hip SystemThere is a potential design issue with the Z1 Offset Stem Inserter leading to fracturing of the distal tip.
-
Class IIofficial
Zimmer, Inc. recalls NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/CIt was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.
-
Class IIofficial
Zimmer, Inc. recalls NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/EIt was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.
-
Class IIofficial
Zimmer, Inc. recalls Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowingA Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery
-
Class IIofficial
Zimmer GmbH recalls Biolox Option Taper SleeveMislabeled: Incorrect sleeve in the packaging. The outer packaging is labeled as a standard neck sleeve, however, the product within the box is a -6mm neck sleeve, and vice versa. The neck of the sleeve can be identified by the device etching as either STD or -6
-
Class IIofficial
Zimmer GmbH recalls Biolox Option Taper SleeveMislabeled: Incorrect sleeve in the packaging. The outer packaging is labeled as a standard neck sleeve, however, the product within the box is a -6mm neck sleeve, and vice versa. The neck of the sleeve can be identified by the device etching as either STD or -6
-
Class IIofficial
Zimmer, Inc. recalls CPT Hip Joint Metal/Polymer Semi-Constrained Cemented ProsthesisAffected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.
-
Class IIofficial
Zimmer, Inc. recalls NexGen CR-Flex Femoral ComponentOut of specification violation of devices that results in a gap existing within the fixture that presses the fiber metal pad to the substrate, causing the pad to not fully bond to the substrate.
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
-
Class IIofficial
Zimmer, Inc. recalls ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular PlateThread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.