Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips Medical Systems Nederland B.v. recalls Azurion 5 M20 System Model Numbers: (1)722228

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Products
  • Azurion 5 M20 System Model Numbers: (1)722228
Where it was sold

Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, FrenchPolynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.

Nationwide

Units affected
922 units (101 US, 821 OUS)
About this recall
Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X