Boston Scientific Corporation recalls Obsidio Conformable Embolic
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.
- Products
-
- Obsidio Conformable Embolic
- Where it was sold
-
US
Nationwide
- Barcodes
0019150603933200191506043124- Units affected
- 985 units
- About this recall
- Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010