Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Calyxo, Inc. recalls CVAC Aspiration System

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.
Products
  • CVAC Aspiration System
Company
Calyxo, Inc.
Where it was sold

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.

Nationwide

Barcodes
00860005357710
Units affected
11,246
About this recall
CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.