Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Zimmer Orthopedics Manfacturing Limited-galway recalls Comprehensive Shoulder Stem

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
Products
  • Comprehensive Shoulder Stem
Company
Zimmer Orthopedics Manfacturing Limited-galway
Where it was sold

US distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.

Nationwide

Barcodes
00887868267550008878682675670088786826757400887868267741
Units affected
90 US; 136 OUS
About this recall
Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627