Zimmer Orthopedics Manfacturing Limited-galway recalls Comprehensive Shoulder Stem
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
- Products
-
- Comprehensive Shoulder Stem
- Company
- Zimmer Orthopedics Manfacturing Limited-galway
- Where it was sold
-
US distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.
Nationwide
- Barcodes
00887868267550008878682675670088786826757400887868267741- Units affected
- 90 US; 136 OUS
- About this recall
- Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627