Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

MEDLINE INDUSTRIES, LP - Northfield recalls Centurion hook kits labeled as: a) STERILE SKIN HOOK FINE (EA3975F)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
Products
  • Centurion hook kits labeled as: a) STERILE SKIN HOOK FINE (EA3975F)
Where it was sold

Worldwide distribution - US Nationwide and the countries of Panama, Canada.

Nationwide

Barcodes
60653160052154606531600618803065316019670300653160317220
Units affected
1520 units
About this recall
Centurion hook kits labeled as: a) STERILE SKIN HOOK FINE (EA3975F), Product Code 66660; b) STERILE SMALL 2 PRONG SKIN HOOK (25754), Product Code 67105; c) STERILE NEWS TRACH HOOK(83300), Product Code 67810; d) STERILE LILLIE EAR HOOK DELICATE 7", Product Code I68655