FDAMedical devicesRecalled Status: Ongoing

Covidien, LP recalls Covidien Signia" Small Diameter Curved Tip Intelligent Reload

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
Products
  • Covidien Signia" Small Diameter Curved Tip Intelligent Reload
Where it was sold

Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.

Nationwide

Units affected
3,023 units
About this recall
Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT