Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Wright Medical Technology, Inc. recalls stryker Infinity Resection Adjustment Block

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.
Products
  • stryker Infinity Resection Adjustment Block
Where it was sold

US: TX, NE, and AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, FRANCE, GERMANY, ITALY, SPAIN, SWEDEN, SWITZERLAND, UK

States: Nebraska, Texas

Barcodes
00889797004008
Units affected
47 units
About this recall
stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty