Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
- Products
-
- CHARLOTTE CLAW Screw
- Where it was sold
-
US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
States: California, Colorado, Florida, Idaho, Michigan, New Jersey, Oklahoma, Pennsylvania, Tennessee, Texas, Wisconsin
- Barcodes
00840420109705- Units affected
- 34
- About this recall
- CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System