Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Navinetics Inc recalls NaviNetics Skull Anchor Key and Drill Kit

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.
Products
  • NaviNetics Skull Anchor Key and Drill Kit
Company
Navinetics Inc
Where it was sold

US Nationwide distribution in the states of FL, MN, CO.

States: Colorado, Florida, Minnesota

Barcodes
10850023620276
Units affected
20 kits
About this recall
NaviNetics Skull Anchor Key and Drill Kit, REF NN1215