Navinetics Inc recalls NaviNetics Skull Anchor Key and Drill Kit
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.
- Products
-
- NaviNetics Skull Anchor Key and Drill Kit
- Company
- Navinetics Inc
- Where it was sold
-
US Nationwide distribution in the states of FL, MN, CO.
States: Colorado, Florida, Minnesota
- Barcodes
10850023620276- Units affected
- 20 kits
- About this recall
- NaviNetics Skull Anchor Key and Drill Kit, REF NN1215