Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Staar Surgical AG recalls EVO+ Visian Toric Implantable Collamer Lens (TICL)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
Products
  • EVO+ Visian Toric Implantable Collamer Lens (TICL)
Where it was sold

International distribution to the countries of India, Iran, Korea, Saudi Arabia.

Barcodes
00840311304363
Units affected
7
About this recall
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5