Staar Surgical AG recalls EVO+ Visian Toric Implantable Collamer Lens (TICL)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
- Products
-
- EVO+ Visian Toric Implantable Collamer Lens (TICL)
- Company
- Staar Surgical AG
- Where it was sold
-
International distribution to the countries of India, Iran, Korea, Saudi Arabia.
- Barcodes
00840311304363- Units affected
- 7
- About this recall
- EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5