Staar Surgical AG recalls Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses.
- Products
-
- Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL
- Company
- Staar Surgical AG
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Austria, Australia, Belgium, Canada, Switzerland, Czech Republic, Germany Spain, Finland, France, Great Britain, Hong Kong, Ireland, Israel, India, Iran, Italy, Jordan, Korea, Kuwait, Kazakhstan, Morocco, Malaysia, Netherlands, Oman, Philippines, Poland, Saudi Arabia, Sweden, Singapore, Thailand, Turkey, Taiwan, Ukraine, USA, Vietnam.
Nationwide
- Barcodes
00840311346288- Units affected
- 543
- About this recall
- Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No