Pyrames Inc recalls Boppli: Bedside Device Kit
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.
- Products
-
- Boppli: Bedside Device Kit
- Company
- Pyrames Inc
- Where it was sold
-
US distribution to states of: MA and CA
States: California, Massachusetts
- Units affected
- 35
- About this recall
- Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007