Pyrames Inc recalls Boppli: Bedside Device Kit

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.
Products
  • Boppli: Bedside Device Kit
Company
Pyrames Inc
Where it was sold

US distribution to states of: MA and CA

States: California, Massachusetts

Units affected
35
About this recall
Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007