Boston Scientific Corporation recalls EndoVive 20Fr Push Standard PEG Kit w/ENFit (Box 2)-Indicated for enteral nutrition directly into the stomach
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration
- Products
-
- EndoVive 20Fr Push Standard PEG Kit w/ENFit (Box 2)-Indicated for enteral nutrition directly into the stomach
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of Japan.
Nationwide
- Barcodes
08714729880493- Units affected
- 572 units
- About this recall
- EndoVive 20Fr Push Standard PEG Kit w/ENFit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00509051