Boston Scientific Corporation recalls EndoVive 20Fr Push Standard PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration
Products
  • EndoVive 20Fr Push Standard PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both
Where it was sold

Worldwide distribution - US Nationwide and the country of Japan.

Nationwide

Barcodes
08714729285205
Units affected
1475 units
About this recall
EndoVive 20Fr Push Standard PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00568211