Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Intuitive Surgical, Inc. recalls Patient Cart

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.
Products
  • Patient Cart
Where it was sold

US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea

States: Alabama, California, Colorado, Florida, Kansas, Kentucky, Maryland, Mississippi, North Carolina, New Hampshire, New Jersey, New York, Ohio, South Carolina, South Dakota, Tennessee, Texas, Virginia, Washington, Wisconsin, West Virginia

Units affected
53
About this recall
Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098