FDAMedical devicesRecalled Status: Ongoing

Tandem Diabetes Care, Inc. recalls t:slim X2 Insulin Pump with Interoperable Technology

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.
Products
  • t:slim X2 Insulin Pump with Interoperable Technology
Where it was sold

US: AZ, CA, CO, CT, FL, HI, IA, ID, IL, IN, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, PA, TX, UT, VA, WA, WI OUS: N/A

States: Arizona, California, Colorado, Connecticut, Florida, Hawaii, Iowa, Idaho, Illinois, Indiana, Maryland, Maine, Michigan, Missouri, Mississippi, North Carolina, North Dakota, Nebraska, New Jersey, New Mexico, Nevada, New York, Ohio, Pennsylvania, Texas, Utah, Virginia, Washington, Wisconsin

Barcodes
00389152407012
Units affected
63 units
About this recall
t:slim X2 Insulin Pump with Interoperable Technology