Medtronic Neuromodulation recalls Stimulation RC Clinician Programmer Application

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is a software issue that can permanently disable communication with an implantable neurostimulator (INS) when a specific intraoperative programming sequence is used.
Products
  • Stimulation RC Clinician Programmer Application
Where it was sold

Worldwide Distribution: US (Nationwide) and OUS (Global) to countries of: Austria, Belgium, Bulgaria, Canary Islands, Cyprus, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Japan, Luxembourg, Malta, Netherlands, Northern Ireland, Norway, Poland, Portugal, Puerto Rico, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Nationwide

Barcodes
00763000371463
Units affected
1493 devices
About this recall
Stimulation RC Clinician Programmer Application, Model A71400, when used with the Inceptiv (Model 977119) and Intellis Pro (Model 977118) neurostimulators