Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Actim Oy recalls Actim PROM dipstik

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Interference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.
Products
  • Actim PROM dipstik
Company
Actim Oy
Where it was sold

US Nationwide distribution in the state of Connecticut.

States: Connecticut

Barcodes
06430030990608
Units affected
6471 kits
About this recall
Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane